Sterilizers using high temperature steam are by far the most used due to their excellent performance.
AF series filters use high-performance filter media providing 100% sterilization efficiency and ensuring the air entering the autoclave is of the highest quality.
They remove solid and biological particles down to 0.01 micron. Their efficiency is higher than that of conventional HEPA filters and has been validated according to standard BS 3928 (Sodium flame test) by a penetration rate of less than 0.001%
AF filters are designed to be reused and are therefore guaranteed for 50 steam sterilization cycles at 134 ° C.
2 models are available to meet the majority of autoclaves available on the market. For use in hospitals/laboratories.
Sterilizers using high-temperature steam are by far the most used due to their excellent performance.
AF series filters use high-performance filter media providing 100% sterilization efficiency and ensuring the air entering the autoclave is of the highest quality.
They remove solid and biological particles down to 0.01 micron. Their efficiency is higher than that of conventional HEPA filters and has been validated according to standard BS 3928 (Sodium flame test) by a penetration rate of less than 0.001%
AF filters are designed to be reused and are therefore guaranteed for 50 steam sterilization cycles at 134 ° C.
2 models are available to meet the majority of autoclaves available on the market.
This product is no longer in stock
Availability date:
* Select the generic if you don't know which model to choose. The sales department will contact you to determine the appropriate model.
(*) Flow rates indicated at atmospheric pressure (20 ° C and 1 Bar absolute).
(1) Construction: Steel, zinc plated.
(2) AF series filters do not eliminate toxic gases.
(3) Connections: Parallel BSP threaded according to ISO 7/1.
Thanks to their excellent performance, autoclaves using high temperature steam are by far the most used to sterilize medical or laboratory equipment.
AF series autoclave filters use a high-performance filter medium that provides 100% sterilization efficiency and guarantees the operator that the air entering the autoclave is of the highest purity.
Manufactured to the highest quality standards, AF filters remove solid particles and microorganisms down to 0.01 micron.
Their effectiveness exceeds that of conventional HEPA filters and has been validated by the flame sodium procedure, according to BS 3928, with a penetration rate of less than 0.001%.
The speed of passage through the media is controlled by a calibrated orifice, according to BS EN 285 standard, and the integrity of each filter is tested before shipment.
Particle filtration | 0,01 Micron |
Penetration at 0,3 micron | < 0,001 % |
Sterilization temperature | 134°C |
Number of guaranteed sterilization cycles | 50s |