Manufacturer | Partenair |
Categories | Compressed air treatment, Compressed air filters |
Document to download | |
C110_filtre_MS |
A range of filters with sterilizable filter elements whose integrity is 100% tested before shipment.
The filter elements, with stainless steel supports and cups, are guaranteed for 100 sterilization cycles at a temperature of 120 ° C.
These filters have been designed for filtration efficiency exceeding the requirements of the "Health technical memorandum 02-01" for medical gas distribution systems.
The range consists of 16 models covering flow rates up to 2550 cubic meters per hour under a pressure of 7 bars.
All filters are delivered complete with their filter element.
This product is no longer in stock
Availability date:
* Select the generic if you don't know which model to choose. The sales department will contact you to determine the appropriate model.
(*) Flow rates indicated at atmospheric pressure (20 ° C and 1 Bar absolute).
(**) As specified by standard HTM 02-01 for medical gas networks.
(1) Connections: parallel BSP according to ISO 7/1.
(2) MS filters must only be installed on a dry, filtered and deoiled air network.
The MS series sterile filters have been developed by WALKER filtration to meet the requirements of the English Health Memorandum HTM 02-01 for medical fluid pipes.
Equipped with sterilizable filter elements, with stainless steel cups, these filters offer performances> 99.9999% at 0.01µ measured by the D.O.P test. (Dioctylphthalate).
The filter elements are designed for a non-stop service life of 1000 hours or 6 months, whichever comes first. They can undergo 100 autoclave sterilization cycles at a temperature of 134 ° C. All items undergo an integrity test and are sterilized (supporting certificate) before shipment.
These filters, intended to retain bacteria and other microorganisms, must be installed as close as possible to use on a dry compressed air line (dew point <= - 20 ° C recommended), filtered and de-oiled.
Particle filtration | 0,01 µ |
Effectiveness of the D.O.P. ** | > 99,9999 % |
Maximum operating temperature (recommended) | 120°C (50°C) |
Own initial pressure drop (replacement) | 100 mbars (400 mbars) |
Replacement of the element (at the first term) | 1000 hours ou 6 months |
Sterilization of the filter element/number of cycles | 134°C / 100 cycles. |
Maximum operating pressure 16 bar | 16 bars |
Differential pressure indicator (A038 to A058) | Pop-up |
Differential pressure indicator (A059 to A390) | Pressure gauge |
Filter elements cups | Stainless steel |
Correction factor according to operating pressure
Pressure | Bars | 4 | 5 | 6 | 7 | 8 | 10 | 12 | 14 | 16 |
Factor | x | 0,76 | 0,84 | 0,92 | 1 | 1,07 | 1,19 | 1,31 | 1,41 | 1,51 |